CVE-2018-10596 Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints

설명

Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

영향 받는 제품 제품 2개

  • Medtronic

    2090 CareLink Programmer

    영향 받는 버전

    • 모든 버전

    해결 버전

    해결된 버전 정보가 없습니다.

  • Medtronic

    29901 Encore Programmer

    영향 받는 버전

    • 모든 버전

    해결 버전

    해결된 버전 정보가 없습니다.