CVE-2021-37164 A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel ope…
설명
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.
영향 받는 제품 제품 1개
-
NVD 기준 해결 버전 있음
swisslog-healthcare
hmi-3 control panel firmware
영향 받는 버전
- 7.2.5.7 미만
해결 버전
- 7.2.5.7 이상
참고 자료 4건
- armis.com https://www.armis.com/PwnedPiper (새 창)
- swisslog-healthcare.com https://www.swisslog-healthcare.com (새 창)
- swisslog-healthcare.com https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 (새 창)