kcve.ai취약점 정보 · 내 자산 확인

CVE-2021-37165 A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel ope…

설명

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When a message is sent to the HMI TCP socket, it is forwarded to the hmiProcessMsg function through the pendingQ, and may lead to remote code execution.

영향 받는 제품 제품 1개

  • swisslog-healthcare

    hmi-3 control panel firmware

    NVD 기준 해결 버전 있음

    영향 받는 버전

    • 7.2.5.7 미만

    해결 버전

    • 7.2.5.7 이상