kcve.ai취약점 정보 · 내 자산 확인

CVE-2022-38100 Contec Health CMS8000

설명

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.

영향 받는 제품 제품 1개

  • Contec Health

    CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor

    영향 받는 버전

    • 모든 버전

    해결 버전

    해결된 버전 정보가 없습니다.

참고 자료 1건

‹ 목록

점검 지원 범위 문의

도입 문의