CVE-2023-0888 Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
설명
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
영향 받는 제품 제품 1개
-
B. Braun
Space Battery Pack SP with Wi-Fi
영향 받는 버전
- 053L000092 이하
- 054U000092 이하
해결 버전
해결된 버전 정보가 없습니다.