CVE-2024-48966 Life2000 service tools for test and calibration do not support user authentication

설명

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

영향 받는 제품 제품 1개

  • Baxter

    Life2000 Ventilation System

    영향 받는 버전

    • 06.08.00.00 이하

    해결 버전

    해결된 버전 정보가 없습니다.