kcve.ai취약점 정보 · 내 자산 확인

CVE-2025-0626 Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

설명

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

영향 받는 제품 제품 1개

  • Contec Health

    CMS8000 Patient Monitor

    영향 받는 버전

    • 모든 버전

    해결 버전

    해결된 버전 정보가 없습니다.

참고 자료 2건

‹ 목록

점검 지원 범위 문의

도입 문의