CVE-2025-14096 Credential Disclosure vulnerability in Radiometer Products

설명

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

영향 받는 제품 제품 3개

  • Radiometer Medical Aps

    ABL90 FLEX and ABL90 FLEX PLUS Analyzers

    영향 받는 버전

    • 버전 정보 확인 필요

    해결 버전

    해결된 버전 정보가 없습니다.

  • Radiometer Medical Aps

    AQT90 FLEX Analyzers

    영향 받는 버전

    • 버전 정보 확인 필요

    해결 버전

    해결된 버전 정보가 없습니다.

  • Radiometer Medical Aps

    ABL800 BASIC and ABL800 FLEX analyzers

    영향 받는 버전

    • 버전 정보 확인 필요

    해결 버전

    해결된 버전 정보가 없습니다.