CVE-2026-66875 Mira Hormone Monitor, Mira Android App Missing authentication for critical function
설명
In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.
영향 받는 제품 제품 2개
-
해결 버전 있음
Quanovate Tech Inc. (operating as Mira / Mira Care)
Mira Firmware
영향 받는 버전
- 1.7.1.47
해결 버전
- 01.07.01.53
-
해결 버전 있음
Quanovate Tech Inc. (operating as Mira / Mira Care)
Mira Android App
영향 받는 버전
- 4.5.15.4
해결 버전
- 4.5.18